Microlife Device Validation Table

 Download Microlife Publication as PDF here.

 Download as PDF here.

Upper arm Automatic
Device AAMI BHS ESH Circumstance Reference
Microlife BP 3AS1-2 Pass At rest 1
Microlife WatchBP O3 (3MZ0) Pass At rest 2
Microlife WatchBP O3 (3MZ1) Pass At rest 3
Microlife WatchBP Office Pass At rest 4
Microlife WatchBP Office ABI A/A Pass At rest 5
Pass L-XL size cuff 4,6
Accuracy in assessing ABI 7
Accuracy in detecting Atrial Fibrillation 8,9
Microlife WatchBP Office AFIB Pass WatchBP Office Equivalence 4*
Accuracy in detecting Atrial Fibrillation 8, 9
Microlife WatchBP Office Target Pass BP A100 Equivalence 10*
 
Self-measurement devices
Device AAMI BHS ESH Circumstance Reference
Microlife BP 3AC1-1 Pass At rest 11
Pass A/A Pregnancy; Abstract 12
Pass B/A Pre-eclampsia; Abstract 12
Microlife BP 3AC1-1 PC Pass BP 3AC1-1 Equivalence 11*
Pass ESRD population 13
Microlife BP 3AC1-2 Pass BP 3AC1-1 Equivalence 11*
Microlife BP 3AG1 A/A BP 3BT0-A Equivalence 14*
Microlife BP 3BTO-1 A/A BP 3BT0-A Equivalence 14*
Microlife BP 3BTO-A A/A Small recruitment violation 14
Pass A/B Normotensive Pregnancy 15
Pass B/B Non-proteinuric HBP 15
Pass A/B Pre-Eclampsia 15
Microlife BP 3BTO-A(2) A/A BP 3BT0-A Equivalence 15*
Microlife BP 3BTO-AP A/A BP 3BT0-A Equivalence 14*
Microlife BP A1 Easy (BP3GR1-1P) A/A Pass BP A100 Plus and BP 3BT0-A Equivalences 16
Pass A/B   Normotensive Pregnancy 10
Pass B/B   Non-proteinuric HBP 10
Pass A/B   Pre-Eclampsia 17
Microlife BP A2 Basic (BP3GQ1-3P) A/A Pass BP A100 Plus and BP 3BT0-A Equivalences 16
Pass A/B   Normotensive Pregnancy 10
Pass B/B   Non-proteinuric HBP 10
Pass A/B   Pre-Eclampsia 17
Microlife BP A6 PC (BP3GU1-8Y) A/A Pass BP A100 Plus and BP 3BT0-A Equivalences 16
Pass A/B   Normotensive Pregnancy 10
Pass B/B   Non-proteinuric HBP 10
Pass A/B   Pre-Eclampsia 17
Microlife BP A100 Pass BP A100 Plus Equivalence 16
Pass With M-L Soft Cuff 10
Pass With M-L Rigid Conical Cuff 17
Microlife BP A100 Plus Pass At rest 10
Pass At rest 17
Microlife Exact BP (BP3MD1-3) Pass BP A100 Equivalence 16
Microlife RM 100 A/A BP 3BT0-A Equivalence 14*
Microlife WatchBP Home (BP3MX1-1) Pass At rest 18
Pass A/A Pregnancy 19
Pass A/A Pre-eclampsia 19
Microlife WatchBP Home A (BP3MX1-3) Pass Watch BP Home Equivalence 18,19*
Accuracy in detecting Atrial Fibrillation 8,9
Microlife WatchBP Home N (BP3MX1-4) Pass Watch BP Home Equivalence 18,19*
Microlife WatchBP Home S (BP3MX1-5) Pass Watch BP Home Equivalence 18,19*
Accuracy in detecting Atrial Fibrillation 8,9
Microlife WatchBP O3 (3MZ0) Pass At rest 2
Microlife WatchBP O3 (3MZ1) Pass At rest 3
 
Wrist devices
Device AAMI BHS ESH Circumstance Reference
Microlife BP W100 Pass A/A Pass Supine position 20
Microlife BP W200-1 Pass MAM averaging and supine position 21

* Indicates that the same validation applies to the device because of equivalence


Sources:

BHS - British Hypertension Society
dableducational Trust



Validation Protocols

 
BHS - British Hypertension Society
             
BHS A/A            
BHS A/A - BHS Protocol Microlife Blood Pressure Monitors are successfully validated according to the protocol of the BHS (British Hypertension Society). Microlife's upper automatic arm devices are validated with the highest possible grade of A/A.  
   
BHS A/B
 
BHS A/B  
             
             

 
European Society of Hypertension
 
             
ESH Protocol
 
Microlife Blood Pressure Monitors are successfully validated according to the International Protocol of the ESH (European Society of Hypertension).  
   
   

 
Gütesiegel
Guetesiegel It is the aim of the German Hypertension Society to improve measurement accuracy, by testing and awarding Blood Pressure Monitoring devices that fulfill its own test criteria following the strict protocol of the German Hypertension Society.  



Publications


1.
de Greeff A, Nathan H, Stafford N, Liu B, Shennan AH. Development of an accurate oscillometric blood pressure device for low resource settings. Blood Press Monit 2008;13:342-8.
2.
Stergiou GS, Tzamouranis D, Nasothimiou EG, Karpettas N, Protogerou A. Are there really differences between home and daytime ambulatory blood pressure? Comparison using a novel dual-mode ambulatory and home monitor. J Hum Hypertens 2010;24:207-12.
3.
Ragazzo F, Saladini F, Palatini P. Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol. Blood Press Monit 2010;15:59-62.
4.
Stergiou GS, Tzamouranis D, Protogerou A, Nasothimiou E, Kapralos C. Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol. Blood Press Monit 2008;13:299-303.
5.
Saladini F, Benetti E, Masiero S, Palatini P. Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 2002 European Society of Hypertension protocol and the British Hypertension Society protocol. Blood Press Monit 2011;16:258-61.
6.
Masiero S, Saladini F, Benetti E, Palatini P. Accuracy of the Microlife large-extra large-sized cuff (32-52 cm) coupled to an automatic oscillometric device. Blood Press Monit 2011;16:99-102.
7.
Kollias A, Xilomenos A, Protogerou A, Dimakakos E, Stergiou GS. Automated determination of the ankle-brachial index using an oscillometric blood pressure monitor: validation vs. Doppler measurement and cardiovascular risk factor profile. Hypertens Res 2011;34:825-30.
8.
Wiesel J, Fitzig L, Herschman Y, Messineo FC. Detection of atrial fibrillation using a modified microlife blood pressure monitor. Am J Hypertens 2009;22:848-52.
9.
Stergiou GS, Karpettas N, Protogerou A, Nasothimiou EG, Kyriakidis M. Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation. J Hum Hypertens 2009;23:654-8.
10.
Stergiou GS, Giovas PP, Neofytou MS, Adamopoulos DN. Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol. Blood Press Monit 2006;11:157-60.
11.
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the SEINEX SE-9400 and the Microlife BP 3AC1-1. Blood Press Monit 2005;10:325-31.
12.
de Greeff A, Reggiori F, Anthony J, Shennan A. The Microlife 3AC1: An accurate blood pressure measurement device in pregnancy and pre-eclampsia. J Hypertens 2006;24.
13.
Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG. Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy. Blood Press Monit 2007;12:227-32.
14.
Cuckson AC, Reinders A, Shabeeh H, Shennan AH. Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. Blood Press Monit 2002;7:319-24.
15.
Reinders A, Cuckson AC, Lee JT, Shennan AH. An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A. BJOG 2005;112:915-20.
16.
Bonso E, Dorigatti F, Palatini P. Accuracy of the BP A100 blood pressure measuring device coupled with a single cuff with standard-size bladder over a wide range of arm circumferences. Blood Press Monit 2009;14:216-9.
17.
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension. Vasc Health Risk Manag 2007;3:389-400.
18.
Stergiou GS, Giovas PP, Gkinos CP, Patouras JD. Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol. Blood Press Monit 2007;12:185-8.
19.
Chung Y, de Greeff A, Shennan A. Validation and compliance of a home monitoring device in pregnancy: microlife WatchBP home. Hypertens Pregnancy 2009;28:348-59.
20.
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of Microlife BP W100 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press Monit 2009;14:41-4.
21.
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of the Microlife BP W200-1 wrist device for blood pressure measurement. Blood Press Monit 2008;13:295-8.