Microlife Microlife
  • Prodotti Domiciliari
    • BP A6 Advanced Easy_red_logo Pressione arteriosa
    • trasp_NON CONTACT Febbre
    • trasp_NEB NANO_half-acc_PRINT_LOcopy1 Terapia respiratoria
    • trasp_WS 80n_half Controllo del peso
    • Software_downloads_allg Software
  • Prodotti Professionali
    • WatchBP Office ABI_front_cuff WatchBP Office
    • WatchBP O3 Ambulatory AFIB_front_cuff WatchBP 03
    • WatchBP Home A_front_cuff WatchBP Home
    • Cuff-Ankle Bracciali e Accessori
    • Software Software
  • Tecnologie
  • Magazine
  • Azienda
    • aboutus Chi siamo
    • Events Eventi e News
    • contact Contatti
    • History Storia
ITALIA
Italiano
Europe
  • EUROPE - English
  • CZECHIA - Čeština
  • DANMARK - Dansk
  • DEUTSCHLAND - Deutsch
  • ESPAÑA - Español
  • FRANCE - Français
  • HRVATSKA - Hrvatski
  • ITALIA - Italiano
  • LATVIJA - Latviešu
  • LIETUVA - Lietuvių
  • MAGYARORSZÁG - Magyar
  • NEDERLAND - Nederlands
  • ÖSTERREICH - Deutsch
  • POLSKA - Polski
  • PORTUGAL - Português
  • SCHWEIZ - Deutsch
  • SLOVENSKO - Slovenčina
  • SLOVENIJA - Slovenščina
  • SUOMI - Suomi
  • SCHWEIZ - Français
  • UK - English
  • ΕΛΛΆΔΑ - Ελληνικά
  • БЪЛГАРИЯ - Български
  • СРБИЈА - Српски
  • УКРАЇНА - Українська
  • TÜRKIYE - Türkçe
  • РОССИЯ - Русский
  • БЕЛАРУСЬ - Русский
Asia
  • ASIA PACIFIC - English
  • VIỆT NAM - Tiếng Việt
  • ประเทศไทย - ไทย
  • 中国 - 中文
  • 台灣 - 中文
Middle East and North Africa
  • MENA - Arabic عربى
USA, Latin America
  • US - English
  • AMÉRICA DEL SUR - Español
  • COLOMBIA - Español
  • MÉXICO - Español
Menu
Microlife_PAD_WristDevice_Header Flow

Validation of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors in a general adult population according to the European Society of Hypertension and the ANSI/AAMI/ISO 81060-2:2013 protocols

Home > Prodotti Professionali
Share

Sen Bing, Kang Chen, Hong Hou, Weijuan Zhang, Linyi Li, Jiao Wei, Chang Shu and Yi Wan, 2015

Objective

This study aimed to determine the accuracy of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors according to the European Society of Hypertension International Protocol revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols.

Methods

The devices were assessed on 33 participants according to the European Society of Hypertension requirements and were then tested on 85 participants according to the ANSI/AAMI/ISO 81060-2:2013 criteria. Procedures and data analysis were carried out following protocol guidelines precisely.

Results

The Microlife BP A200 Comfort and W2 Slim devices passed the criteria of the European Society of Hypertension International Protocol revision 2010 for both systolic blood pressure and diastolic blood pressure. The devices also fulfilled the ANSI/AAMI/ISO 81060-2:2013 criteria, with mean differences of SBP and DPB between the devices and observers of 0.38 ±5.12 and 0.28± 4.29 mmHg for the BP A200 Comfort and 1.01± 6.80 and 0.34 ± 5.62 mmHg for the W2 Slim,  respectively.

Conclusion

The Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors fulfilled the European Society of Hypertension revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols, and can be recommended for self-measurement in the general population.

Click here to download abstract as PDF